As a rule, the pH of the mobile phase buffer must be between the pI (isoelectric point) or pKa (acid dissociation constant) of the charged molecule and the pKa of the charged group on the solid support. For example, in cation exchange chromatography, using a functional group on the solid support with a pKa of 1.2, a sample molecule with a pI of 8.2 may be run in a mobile phase buffer of pH 6.0. In anion exchange chromatography a molecule with a pI of 6.8 may be run in a mobile phase buffer at pH 8.0 when the pKa of the solid support is 10.3.