Characteristics of Heparin Sepharose and Capto Heparin chromatography media</p>
| Ligand density (mg/ml) | Compositionp | pH stability1 | Average particle size (µm) | |
|---|---|---|---|---|
| Heparin Sepharose High Performance | 10 | Heparin coupled to Sepharose High Performance by reductive amination to give a stable attachment even in alkaline conditions. | Short term: 5 to 10 Long term: 5 to 10 |
34 |
| Heparin Sepharose 6 Fast Flow | 5.0 | Heparin coupled to Sepharose 6 Fast Flow by reductive amination to give a stable attachment even in alkaline conditions. | Short term: 4 to 13 Long term: 4 to 12 |
90 |
| Capto Heparin | 1.8 | Heparin coupled to Capto | Short term: 4 to 13 Long term: 4 to 12 |
90 |
1 Short term refers to the pH interval for regeneration, cleaning-in-place, and sanitization procedures. Long term refers to the pH interval over which the medium is stable over a long period of time without adverse effects on its subsequent chromatographic performance.
HiTrap Heparin HP and Heparin Sepharose 6 Fast Flow
[edit]Heparin Binding buffer: 20 mM Tris-HCl, pH 8.0 or 10 mM sodium phosphate, pH 7.0
Heparin Elution buffer: 20 mM Tris-HCl, 1 to 2 M NaCl, pH 8.0 or 10 mM sodium phosphate, 1 to 2 M NaCl, pH 7.0
Equilibrate the column with 10 CV of binding buffer.
Load the sample.
Wash with 5 to 10 CV of Heparin binding buffer or until no material appears in the eluent (monitored by UV absorption at A280 nm).
Elute with 5 to 10 CV of Heparin elution buffer using a continuous or step gradient from 0% to 100% elution buffer.
Example
[edit]After Purification from Ni, Desalt using buffer: 20 mM Tris-HCl, pH 8, 5% (v/v) glycerol, 10 mM 2-mercaptoethanol, 0.1 M NaCl
Then load 35 ml to HiTrap Heparin HP 5 ml column.
Binding buffer: 20 mM Tris-HCl, pH 8, 5% (v/v) glycerol, 10 mM 2-mercaptoethanol, 0.1 M NaCl
Elution buffer: 20 mM Tris-HCl, pH 8, 5% (v/v) glycerol, 10 mM 2-mercaptoethanol, 1 M NaCl
Flow rate: 1 ml/min (30 cm/h)
Gradient: Linear, 0.1 M NaCl to 1 M NaCl in 20 column volumes
Fraction volume: 3 ml
Cleaning
Remove ionically bound proteins by washing with 0.5 CV of 2 M NaCl for 10 to 15 min.
Remove precipitated or denatured proteins by washing with 4 CV of 100 mM NaOH for 1 to 2 h; or 2 CV of 6 M guanidine hydrochloride for 30 to 60 min; or 2 CV of 8 M urea for 30 to 60 min. Remove hydrophobically bound proteins by washing with 4 CV of 0.1% to 0.5% Tween 20 for 1 to 2 h.
Chemical stability
100 mM NaOH (1 w at 20°C), 50 mM sodium acetate, pH 4.0, 4 M NaCl, 8 M urea, 6 M guanidine hydrochloride.
Storage
Wash chromatography media and columns with 50 mM sodium acetate containing 20% ethanol (use approximately 5 CV for packed media) and store at 4°C to 8°C.
Capto Heparin
[edit]Binding buffer: 100 mM Tris-HCl, 10 mM trisodium citrate, 225 mM NaCl, pH 7.4
Wash buffer: 100 mM Tris-HCl, 10 mM trisodium citrate, 330 mM NaCl, pH 7.4
Elution buffer: 100 mM Tris-HCl, 10 mM trisodium citrate, 2 M NaCl, pH 7.4
Equilibrate with 5 CV of binding buffer.
Apply the sample.
Wash step 1: wash with 40 CV of binding buffer.
Wash step 2: wash with 15 CV of wash buffer.
Elute with 9.5 CV of elution buffer.
A flow rate of 0.5 ml/min is recommended for a 1 ml column.
Cleaning
Substances such as denatured proteins that do not elute during regeneration can be removed by cleaning-in-place (CIP) procedures.
A recommended CIP procedure for Capto Heparin is 4 CV of 100 mM NaOH with a contact time of 1 to 2 h.
Storage
Store unused chromatography media at 4ºC to 30ºC in 20% ethanol and 50 mM of sodium acetate.